• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MAESTRO 4000 CONTROLLER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MAESTRO 4000 CONTROLLER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER Back to Search Results
Model Number M00440000
Device Problem Connection Problem (2900)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/01/2019
Event Type  malfunction  
Event Description
It was reported that non-irrigated ablation occurred.A maestro 4000 system was selected for an ablation procedure.However it was noted that the cable that connects the metriq pump to the maestro generator didn't screw in correctly and came loose, this resulted in an ablation with out using irrigation.No patient complications were reported and the procedure was completed with this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAESTRO 4000 CONTROLLER
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
STELLARTECH RESEARCH CORPORATION
560 cottonwood drive
,
milpitas CA 95035
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330,
saint paul, MN 55112
6515826168
MDR Report Key8374741
MDR Text Key137291831
Report Number2134265-2019-01790
Device Sequence Number1
Product Code LPB
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P020025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 02/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00440000
Device Catalogue Number86248
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/01/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Type of Device Usage N
Patient Sequence Number1
-
-