MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD. 801037 ASR TAP SLV ADAP 12/14 -1; ASR HIP SYSTEM : HIP FEMORAL SLEEVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD. 801037 ASR TAP SLV ADAP 12/14 -1; ASR HIP SYSTEM : HIP FEMORAL SLEEVE Back to Search Results
Catalog Number 999890353
Device Problems Biocompatibility (2886); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Host-Tissue Reaction (1297); Calcium Deposits/Calcification (1758); Foreign Body Reaction (1868); Pain (1994); Injury (2348); Deformity/ Disfigurement (2360); Osteolysis (2377); Test Result (2695); No Code Available (3191)
Event Date 02/04/2019
Event Type  Injury  
Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event Description
Litigation alleged the patient suffered pain, permanent physical injuries, disfigurement and exposure to excessive levels of chromium and cobalt as a result of the implanted asr hip.Update ad 04 feb 2019: receipt of der.There were no new allegation stated however, patient has been revised due to failed of the asr xl acetabular implant.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> the asr platform was voluntarily recalled from the market in august 2010, and the asr product codes are now considered inactive.Further investigation of this individual incident will not be undertaken.The investigation regarding the root cause(s) and/or corrective actions was conducted under mdd capa-(b)(4).Ongoing post market surveillance is conducted per our procedures for this product.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Patient code: no code available (3191) used to capture the medical device removal.
 
Event Description
After review of medical records the patient was revised to address pain and elevated metal ions.Operative finding reported heterotopic ossification, metallosis in the capsule, staining of the trunnion femoral component with proximal osteolysis and no bony ingrowth in the acetabular component.Surgical prolong for 45-50 minutes due to obesity and extensive ho.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: corrected: h6(device code).Product complaint (b)(4).Investigation summary : no device associated with this report was received for examination.This hip replacement platform was voluntarily recalled from the market and the product codes are now considered inactive.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASR TAP SLV ADAP 12/14 -1
Type of Device
ASR HIP SYSTEM : HIP FEMORAL SLEEVE
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD. 801037
st. anthony's rd
leeds LS11 8 DT
UK  LS11 8 DT
Manufacturer (Section G)
DEPUY INTERNATIONAL LTD. 8010379
st. anthony's rd
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kara ditty-bovard
1210 ward avenue
west chester, PA 19380-XXXX
6107428552
MDR Report Key8375303
MDR Text Key137302994
Report Number1818910-2019-85842
Device Sequence Number1
Product Code KXA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,Company Representative
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 02/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/28/2014
Device Catalogue Number999890353
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/17/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/18/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ASR ACETABULAR CUPS 52.; ASR UNI FEMORAL IMPL SIZE 46.; CORAIL2 STD SIZE 14.
Patient Outcome(s) Required Intervention;
Patient Age82 YR
Patient SexMale
Patient Weight102 KG
-
-