Device Problems
Use of Device Problem (1670); Material Perforation (2205); Device Dislodged or Dislocated (2923)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); No Information (3190)
|
Event Date 01/31/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The manufacturer's investigation is on-going and further information will be provided once the investigation has been completed.
|
|
Event Description
|
The customer reported that the vaginal tube for the 25, 30 and 25 mm set will not stay locked in place.Unfortunately a patient's vaginal cuff was perforated when the vaginal tube became dislodged.
|
|
Event Description
|
The customer reported that the vaginal tube for the 25, 30 and 25 mm set will not stay locked in place.Unfortunately a patient's vaginal cuff was perforated when the vaginal tube became dislodged.
|
|
Manufacturer Narrative
|
The investigation was completed by conducting a thorough evaluation of the reported information.The intrauterine tube has a diameter of 14 mm which implies that considerable force is required to penetrate/puncture patient tissue.This force most likely occurred when the patient was moved to the imaging room with the applicator fixated in the applicator clamp.The applicator was fixated with the ct/mr applicator clamp directly on the intra uterine tube instead of on the cylinder using the designated adapter (110778 - optional part) as per the user manual.Due to this the iu tube can be extended from the cylinder if it is not fixated adequately with the screw when the patient moves while being transferred to the image room.The cylinder and screws were not returned, it could not be determined if the clamping force was adequate or not.It is recommended to use the designated clamp adapter to fixate on the cylinder and not on the intra uterine tube when using the ct/mr applicator clamp.Patient movement must be avoided when the applicator is fixated in the clamp.The labelling cautions: to avoid excessive external forces.Therefore if this issue is applicator related, it is considered a use error.
|
|