Catalog Number CAT8TORQ85 |
Device Problems
Device Damaged Prior to Use (2284); Material Deformation (2976)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 01/31/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.
|
|
Event Description
|
During preparation for a medical procedure, an indigo system aspiration catheter 8 (cat8) was found to be ovalized at the distal end upon removal from the packaging.The cat8 was therefore not used in the procedure.The procedure was completed using another cat8.
|
|
Manufacturer Narrative
|
Results: the returned device was ovalized from approximately 2.0 cm to 4.0 cm from the hub.Conclusions: evaluation of the returned cat8 revealed a device with an ovalization on the proximal end.If the device is forcefully gripped or pinched during removal from its packaging, damage such as this may occur.No damage was present on the distal end of the device.The damage in the reported complaint was likely the ovalization found on the proximal end of the device.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
|
Search Alerts/Recalls
|