ARROW INTERNATIONAL INC. EPIDURAL CATHETERIZATION KIT; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number WJ-05401
Device Problem Obstruction of Flow (2423)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/14/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device investigation is pending.Teleflex will continue to monitor and trend related events.
 
Event Description
It was another one that had been in 4 hours and stop working after that; was not able to flush at that point.
 
Manufacturer Narrative
Qn#(b)(4).A lot number was not provided.A device history record review was performed based upon a lot number from sales history data.A device history record review was performed on the epidural catheter with no relevant findings.The customer reported the catheter was blocked.The customer returned one snaplock adapter and one epidural catheter.The components were received connected together.The returned components were visually examined with and without magnification.Visual examination of the returned snaplock adapter revealed that the snaplock adapter typical with no observed defects or anomalies.Visual examination of the returned catheter revealed that the catheter appears typical but used.Adhesive residue is present on the catheter body exterior and biological material can be seen between the inner coils.No other defects or anomalies were observed.A functional flow test was performed on the returned sample per amrq-000017 section 7.8; rev.7.The returned epidural catheter was inserted from the proximal end into the returned snaplock adapter until it bottomed out and the snaplock adapter was closed.The components were confirmed to be secured by tugging gently.The snaplock adapter was connected to the lab leak tester (c05176) and the pressure was increased to 10 psi to establish flow.Water could be seen immediately exiting the distal end of the catheter.The flow rate was measured at 4.6ml/min (c05180), which is within the specification of 1ml/min.No blockages were found.The reported complaint of the catheter being blocked could not be confirmed based on the sample received.A lot number was not provided.A device history record review was performed based upon a lot number from sales history data.A device history record review was performed on the epidural catheter with no evidence to suggest a manufacturing related issue.During functional inspection, the returned components passed a functional flow test and met flow rate specifications.There were no functional issues found with the returned sample.
 
Event Description
It was another one that had been in 4 hours and stop working after that; was not able to flush at that point.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION KIT
Type of Device
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key8386982
MDR Text Key137715178
Report Number1036844-2019-00203
Device Sequence Number1
Product Code CAZ
Combination Product (y/n)N
PMA/PMN Number
K103658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/04/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberWJ-05401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2019
Date Manufacturer Received04/02/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-