• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES, PR SWAN-GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES, PR SWAN-GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION Back to Search Results
Model Number PE074F5
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Appropriate Term/Code Not Available (3191)
Patient Problem Cardiac Perforation (2513)
Event Date 02/13/2019
Event Type  Injury  
Manufacturer Narrative
The product is expected to be returned for analysis; however, it has not yet been received.Upon the return of the product a supplemental report will be sent with the investigation results.The lot number was not provided; therefore, a review of the manufacturing records could not be completed.
 
Event Description
It was reported that a patient experienced cardiac perforation during use of a swan ganz pacing catheter.It was reported that the tip of the catheter sunk into the interventricular septum near the cardiac apex during leadless pacemaker implantation.The severity was determined as ¿not serious¿.The patient status was reported as ¿recovered¿.Patient demographic information requested but unavailable.The causality between the device and the event was reported as unknown.At this time the physician declined to share any further information.
 
Manufacturer Narrative
No product was returned for evaluation; it was discarded at the hospital.Without the return of the product, it is not possible to determine if damages or defects existed on the product, nor could a root cause or potential contributing factors be identified.No further actions will be taken.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised, before deciding to insert or use the catheter, to consider the potential benefits in relation to the possible complications.The techniques for insertion, methods of using the catheter to obtain patient data information, and the occurrence of complications is well described in the literature.With any hemodynamic monitoring, patient parameters can change quickly and dramatically.Patient parameters should correlate with the patient¿s clinical manifestations.Cases of myocardial perforation associated with the use of temporary trans-venous pacing catheters have been reported.Perforation/damage may occur as a result of over-wedging a pulmonary artery catheter or forcing the advancement of the catheter against resistance.Any perforation may cause significant bleeding/ injury.Careful repositioning and withdrawal of the catheter under fluoroscopic control is recommended.It is unknown if user or procedural factors may have contributed to this event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Manufacturer Narrative
Corrected data: f10, h6.Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION
Type of Device
PACING CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
MDR Report Key8392654
MDR Text Key137931772
Report Number2015691-2019-00737
Device Sequence Number1
Product Code LDF
Combination Product (y/n)N
PMA/PMN Number
K813521
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 02/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/05/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberPE074F5
Was Device Available for Evaluation? No
Date Manufacturer Received07/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-