Model Number WA22351A
Device Problems Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/26/2019
Event Type  malfunction  
Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event Description
Olympus was informed that during a therapeutic transurethral resection of the bladder tumor (turbt) procedure, the loop wire at the distal end of the hf resection electrode broke off and fell into the patient¿s bladder.However, no fragment remained inside the patient since it was reportedly retrieved.The intended procedure was successfully completed with a similar device and there was no report about an adverse event or patient injury.
 
Manufacturer Narrative
The suspect medical device was not returned to the manufacturer for evaluation/investigation.Therefore, the manufacturer's evaluation was exclusively performed on the basis of the provided information.The subject hf-resection electrode is a reusable product.According to the provided information, the error occurred on first use.Since there is presently no indication of an increased number of similar cases related to this product, the most likely reason is wear and tear.Another possible cause of this damage is mechanical overload as a result of the application of excessive force.Therefore, this event/incident was attributed to use error.Furthermore, a material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the hf-resection electrode without showing any abnormalities.The case will be closed from olympus side with no further actions but the reported event/incident will be recorded for trending and surveillance purposes.Furthermore, the user will be informed about the investigation results and retrained to correctly use the olympus medical devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, ROLLER, 24-28 FR., 12° AND 30°, FOR TURIS
Type of Device
HF-RESECTION ELECTRODES
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
MDR Report Key8394131
MDR Text Key141483245
Report Number9610773-2019-00053
Device Sequence Number1
Product Code FAS
UDI-Device Identifier04042761030526
UDI-Public04042761030526
Combination Product (y/n)N
PMA/PMN Number
K120418
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 06/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/06/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA22351A
Device Catalogue NumberWA22351A
Device Lot Number1000023302
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/29/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age95 YR
-
-