Model Number 85020 |
Device Problems
Thermal Decomposition of Device (1071); High impedance (1291)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/14/2019 |
Event Type
malfunction
|
Event Description
|
It was reported that catheter charring occurred.During an ablation procedure with a blazer ii ablation catheter, the initial impedance in the iliac vein was very high (230 ohms) and about ten radiofrequency shots were performed at low power (10 watts).Upon removal of the catheter, it's tip was noted to be charred.A second device of the same batch was placed but since the initial impedance was high, no shot was performed.A third device from another batch was used and this was functional.There were no consequences for the patient.
|
|
Manufacturer Narrative
|
The device was returned for analysis.Overall visual revision did not identify failures or evidence that could be lost due to decontamination process.Visual inspection revealed char observed on the tip in distal end.Functional inspection revealed the steering knob and tension control knob functioned properly on both lock and unlock positions.No abnormal resistance was felt when actuating the steering mechanism.Electrical test revealed d02-high impedance message appeared on the maestro 4000 screen, since the distal tip section with char was in touch with the connection fixture, therefore, the section of the tip with char was replaced making the electrical contact with the fixture and running the test without issues.The ablation was verified by using the maestro generator 4000, and the device was found within specifications.
|
|
Event Description
|
It was reported that catheter charring occurred.During an ablation procedure with a blazer ii ablation catheter, the initial impedance in the iliac vein was very high (230 ohms) and about ten radiofrequency shots were performed at low power (10 watts).Upon removal of the catheter, it's tip was noted to be charred.A second device of the same batch was placed but since the initial impedance was high, no shot was performed.A third device from another batch was used and this was functional.There were no consequences for the patient.
|
|
Search Alerts/Recalls
|