The device was returned for analysis.A visual and microscopic examination of the balloon, blade, merkerbands, hypotube and shaft.The balloon noted that the balloon was tightly folded and had not been subjected to positive pressure.No issues were identified with the balloon material that could have contributed to the complaint incident.Blades noted that all blades were present and fully bonded to the balloon material.No damage or any issues were noted with the blades that could have contributed to the complaint incident.No issues were observed with the tip or markerbands.It was noted that the break in the hypotube was approximately 1420mm proximal to the tip.The detached section of hypotube and manifold were not returned for analysis.Multiple kinks were also noted along the length of the hypotube.This type of damage is consistent with excessive force being applied to the delivery system.No issues were noted with the hypotube that could have contributed to the damage identified.An approximate 6mm kinked was noted proximal to the proximal markerband on the saft polymer extrusion.This type of damage is consistent with excessive force being applied to the delivery system.No other issues were identified during the product analysis.
|
Reportable based on new information received on (b)(6) 2019.It was reported that difficulty crossing lesion occurred.The target lesion was located in the left coronary artery.A 15/3.00 flextome cutting balloon was selected for use.During procedure, it was noted that the device could not cross the lesion.The procedure was completed using a different device of the same.No patient complications were reported.However, additional information revealed that the device was returned with a broken hypotube and the manifold completely detached.
|