One uni-directional, curve f, tacticath sensor enabled contact force ablation catheter was received for evaluation.No visual anomalies were noted.Electrical testing of the returned device determined electrodes 1-4 and the magnetic sensor met specifications for acceptable resistance values with no open or short circuits detected.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported display and impedance issues remains unknown.
|