• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEUROPACE, INC. NEUROPACE RNS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEUROPACE, INC. NEUROPACE RNS SYSTEM Back to Search Results
Model Number DL-330-10-K
Device Problems Break (1069); High impedance (1291)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/18/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4) review of the device ecog and lead impedance confirmed that the ecog data are suggestive of a potential lead break.
 
Event Description
A change in the signals for the depth lead connected to port 1 of the rns neurostimulator was noticed on (b)(6) 2017.The treating center reported that the patient had a post-ictal altercation on (b)(6) 2017 in which the patient's head hit the ground.On (b)(6) 2018, detection was disabled on the depth lead connected to port 1.During a routine neurostimulator replacement on (b)(6) 2019, the lead was removed and a new lead was implanted and connected to the neurostimulator without complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Type of Device
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer (Section G)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer Contact
ramona gonis
455 n. bernardo ave.
mountain view, CA 94043
6502382788
MDR Report Key8417451
MDR Text Key138691581
Report Number3004426659-2019-00007
Device Sequence Number1
Product Code PFN
UDI-Device Identifier00855547005045
UDI-Public010085554700504517200508
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/13/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Model NumberDL-330-10-K
Device Catalogue Number1007605
Device Lot Number23840-1-1-1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/18/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other;
Patient Age31 YR
-
-