According to the reporter, the device displayed zero respiratory and heart rate at 16:40 but on the other unit it was displayed 99 spo2 and 150 heart rate.It was also stated that the device was turned off and the sensor was removed after noticing that it was still remaining on the values stated.The beep sound was managed silence in the hospital and so the device was not sounding.The patient death was confirmed at 18:20 but the clock on the device displayed 7:26 at 9:13.A gap was found by 1 hour and 47 minutes.The cause of death was unknown.
|