Model Number CI-1400-02H |
Device Problems
Defective Component (2292); Therapeutic or Diagnostic Output Failure (3023)
|
Patient Problem
Failure of Implant (1924)
|
Event Date 02/14/2019 |
Event Type
malfunction
|
Event Description
|
The recipient reportedly experienced shorted and open electrodes and decreased performance.Programming adjustments were made, however, the issue did not resolve.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
|
|
Manufacturer Narrative
|
The recipient's activation reportedly went well.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The device passed the external visual inspection.The photographic imaging inspection revealed broken electrode wires near the electrode ground ring.System lock was verified.The electrode condition prevented some of the electrical tests from being performed.The device passed some of the electrical tests performed.The device passed the mechanical tests performed.Dissection and the scanning electron microscopy analysis of the electrode revealed evidence of fatigue breaks near the electrode ground ring region.The photographic imaging inspection revealed broken wires in the fantail region.It is believed that these breaks caused the decreased performance.A corrective action was implemented.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|