• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH HLM TUBING SET W/BIOLINE COATING; TUBING, PUMP, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH HLM TUBING SET W/BIOLINE COATING; TUBING, PUMP, CARDIOPULMONARY BYPASS Back to Search Results
Model Number BEQ-HLS 5050 USA#HL
Device Problem Output Problem (3005)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/15/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A follow-up medwatch will be submitted when additional information becomes available.(b)(4).The product was requested for return to the manufacturer for laboratory investigation but the product is already scrapped.The product of the related (b)(4) will be returned for lab investigation.The investigation is still pending.A supplemental medwatch will be submitted after new information has been received.
 
Event Description
They stated that they are supporting a patient and they believe the module component is not performing correctly.They have it maxed out on flow and sweep and its not responding the way they would expect.The device was replaced during treatment.No harm to the patient was reported.(b)(4).
 
Event Description
This complaint relates to tw # (b)(4).Internal reference: (b)(4).
 
Manufacturer Narrative
Maquet medical systems,usa(importer)submits this report on behalf of the legal manufacturer of the device maquet cardiopulmonary ag kehler strasse 31, 76437 rastatt, germany.A follow-up medwatch will be submitted when additional information becomes available.Reference exemption # e2018002.Importer- maquet medical systems usa, 45 barbour pond drive, wayne, nj 07470.Contact person- pedro cruz the product was requested for return to the manufacturer for laboratory investigation but the product is already scrapped.The product of the related tw complaint # (b)(4) was also discarded.Therefore no laboratory investigation could be performed by the manufacturer.A review for similar complaints to be investigated already was performed and no similar complaints were found.Thus the reported failure could not be confirmed.Based on the information available at this time the cause of this failure was determined to not be attributed to attributed to a device related malfunction.Since the reported failure did not contribute to a death or serious injury no corrective action is needed.In addition at this time it cannot be concluded that this is a systemic error.No corrective action is needed.The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HLM TUBING SET W/BIOLINE COATING
Type of Device
TUBING, PUMP, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
MDR Report Key8445297
MDR Text Key139811652
Report Number8010762-2019-00076
Device Sequence Number1
Product Code DWE
Combination Product (y/n)N
PMA/PMN Number
K080592
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 03/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/22/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBEQ-HLS 5050 USA#HL
Device Catalogue Number701052797
Was Device Available for Evaluation? No
Date Manufacturer Received03/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-