Catalog Number 8065977763 |
Device Problems
Break (1069); Scratched Material (3020)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/27/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The product was not returned for analysis.The complaint history and product history records could not be reviewed because the reporting facility did not provide a lot number or any identification traceable to the manufacturing documentation.The product investigation could not identify a root cause.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A facility representative reported that during an intraocular lens (iol) implant procedure, the lens was scratched.The surgery was completed with a back up lens.According to the surgeon's opinion, the injector caused or contributed to the event.Additional information was provided indicating that the iol was damaged during delivery through the cartridge.Two paracentral cracks were visable in the optic after delivery in the eye.The iol was removed and replaced with another.
|
|
Manufacturer Narrative
|
Evaluation summary: the cartridge was not returned for analysis.The lens was returned with extensive damage.The lens was in two pieces typical of a removal.The iol product history records were reviewed and the documentation indicated the product met release criteria.The customer indicated the use of a viscoelastic, which is not qualified for this lens/cartridge combination used.The root cause may be related to a failure to follow the directions for use (dfu).Due to differing material properties, the use of a non-qualified viscoelastic may result in delivery issues and/or damage.Lens damage during delivery typically occurs if the lens is not positioned correctly/folded correctly for advancement; if there is a lack of viscoelastic between the lens and the cartridge lumen; and/or if the handpiece plunger is not positioned correctly at the trailing optic edge.This can allow the lens to fold around the plunger tip making it too large to correctly advance through the narrow tip of the cartridge causing damage.The manufacturer internal reference number is: (b)(4).
|
|
Search Alerts/Recalls
|