MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL) Back to Search Results
Catalog Number 8065977763
Device Problems Break (1069); Scratched Material (3020)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/27/2019
Event Type  malfunction  
Manufacturer Narrative
The product was not returned for analysis.The complaint history and product history records could not be reviewed because the reporting facility did not provide a lot number or any identification traceable to the manufacturing documentation.The product investigation could not identify a root cause.The manufacturer internal reference number is: (b)(4).
 
Event Description
A facility representative reported that during an intraocular lens (iol) implant procedure, the lens was scratched.The surgery was completed with a back up lens.According to the surgeon's opinion, the injector caused or contributed to the event.Additional information was provided indicating that the iol was damaged during delivery through the cartridge.Two paracentral cracks were visable in the optic after delivery in the eye.The iol was removed and replaced with another.
 
Manufacturer Narrative
Evaluation summary: the cartridge was not returned for analysis.The lens was returned with extensive damage.The lens was in two pieces typical of a removal.The iol product history records were reviewed and the documentation indicated the product met release criteria.The customer indicated the use of a viscoelastic, which is not qualified for this lens/cartridge combination used.The root cause may be related to a failure to follow the directions for use (dfu).Due to differing material properties, the use of a non-qualified viscoelastic may result in delivery issues and/or damage.Lens damage during delivery typically occurs if the lens is not positioned correctly/folded correctly for advancement; if there is a lack of viscoelastic between the lens and the cartridge lumen; and/or if the handpiece plunger is not positioned correctly at the trailing optic edge.This can allow the lens to fold around the plunger tip making it too large to correctly advance through the narrow tip of the cartridge causing damage.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D
Type of Device
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
MDR Report Key8451313
MDR Text Key139972607
Report Number1119421-2019-00354
Device Sequence Number1
Product Code MSS
Combination Product (y/n)N
PMA/PMN Number
K063155
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other
Type of Report Initial,Followup
Report Date 06/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/26/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065977763
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2019
Date Manufacturer Received05/23/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
MONARCH III HANDPIECE; PROVISC; SN6AT3.225 DIOPTER
-
-