• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALPHATEC SPINE INC ANTERIOR CERVICAL PLATE ASSEMBLY, 2-LEVEL, 026 MM (TI-6AI-4V ELI, NITINOL); SPINAL INTERVERTEBRAL BODY FIXATION ORTHOSIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALPHATEC SPINE INC ANTERIOR CERVICAL PLATE ASSEMBLY, 2-LEVEL, 026 MM (TI-6AI-4V ELI, NITINOL); SPINAL INTERVERTEBRAL BODY FIXATION ORTHOSIS Back to Search Results
Model Number 71002-026
Device Problem Unintended Movement (3026)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/14/2018
Event Type  malfunction  
Manufacturer Narrative
No evaluation possible at this time.The international customer has not returned the device for evaluation.Upon the receipt of additional information and/or the product in question, a follow-up report will be submitted.
 
Event Description
When placing the plate and inserting the screws for tightening, the secure central golden piece jumped.Fortunately, it was retrieved with no consequences to the patient.Finally, a new plate was used instead.No patient harm reported.
 
Manufacturer Narrative
Visual examination of the returned 2-level plate found that the middle blocking slide had become completely detached from the implant.Witness marks along both the anterior and posterior sides of the plate are present and reveal the location of the instrument to plate during contouring/bending.The plate contains a witness mark on the posterior side directly behind the location of the detached blocking slide and witness marks at both graft windows on the anterior side.The combination of these witness marks confirms the plate bending instrument was positioned inconsistent with ifu requirements while contouring/bending the plate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTERIOR CERVICAL PLATE ASSEMBLY, 2-LEVEL, 026 MM (TI-6AI-4V ELI, NITINOL)
Type of Device
SPINAL INTERVERTEBRAL BODY FIXATION ORTHOSIS
Manufacturer (Section D)
ALPHATEC SPINE INC
5818 el camino real
carlsbad CA 92008
MDR Report Key8451790
MDR Text Key140311347
Report Number2027467-2019-00012
Device Sequence Number1
Product Code KWQ
UDI-Device Identifier00844856066977
UDI-Public(01)00844856066977(10)697688
Combination Product (y/n)N
PMA/PMN Number
K102820
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 02/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/26/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71002-026
Device Catalogue Number71002-026
Device Lot Number697688
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2019
Date Manufacturer Received02/26/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-