ALPHATEC SPINE INC ANTERIOR CERVICAL PLATE ASSEMBLY, 2-LEVEL, 026 MM (TI-6AI-4V ELI, NITINOL); SPINAL INTERVERTEBRAL BODY FIXATION ORTHOSIS
|
Back to Search Results |
|
Model Number 71002-026 |
Device Problem
Unintended Movement (3026)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/14/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
No evaluation possible at this time.The international customer has not returned the device for evaluation.Upon the receipt of additional information and/or the product in question, a follow-up report will be submitted.
|
|
Event Description
|
When placing the plate and inserting the screws for tightening, the secure central golden piece jumped.Fortunately, it was retrieved with no consequences to the patient.Finally, a new plate was used instead.No patient harm reported.
|
|
Manufacturer Narrative
|
Visual examination of the returned 2-level plate found that the middle blocking slide had become completely detached from the implant.Witness marks along both the anterior and posterior sides of the plate are present and reveal the location of the instrument to plate during contouring/bending.The plate contains a witness mark on the posterior side directly behind the location of the detached blocking slide and witness marks at both graft windows on the anterior side.The combination of these witness marks confirms the plate bending instrument was positioned inconsistent with ifu requirements while contouring/bending the plate.
|
|
Search Alerts/Recalls
|
|
|