Qn# (b)(4).The dhr for the returned instrument was reviewed and found completely without any irregularities.This instrument was produced at the tecomet , inc.Kenosha, wi facility as part of a 50 pc.Lot in february of 2018.The returned instrument as received was evaluated and found that the jaws are not misaligned as stated in complaint and this instrument is able to pick-up, retain, close and release multiple clips both with and without the use of silastic test tubing.We are unable to validate this complaint since we the jaws are aligned , and we are unable to locate any non-conforming features on the returned sample part.All 50 instruments from this lot were 100% visually inspected and function tested prior to shipment to the customer as this is a standardized procedure at this facility for this product line.
|