Catalog Number 5MAXACE068 |
Device Problems
Device Damaged Prior to Use (2284); Material Deformation (2976)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 02/05/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.
|
|
Event Description
|
During preparation for a thrombectomy procedure, the physician noticed that the proximal end of penumbra system ace 68 reperfusion catheter (ace68) was kinked upon removal from packaging.The damage to the ace68 was found prior to use.Therefore, the ace68 was not used in the procedure.The procedure was completed using a new ace68.
|
|
Manufacturer Narrative
|
Results: the returned device was kinked at approximately 1.0 cm from the hub under the strain relief.The device was ovalized at approximately 37.0 cm and 101.0 cm from the hub.Conclusions: evaluation of the returned ace68 confirmed a kinked under the strain relief on the proximal end of the catheter.If the device is forcefully mishandled during removal from its packaging, damage such as this may occur.Further investigation revealed ovalizations on the catheter shaft.These damages were likely incidental to the complaint.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
|
Search Alerts/Recalls
|