• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL PUERTO RICO, INC. SJM MASTERS SERIES VALVE EXPANDED CUFF; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL PUERTO RICO, INC. SJM MASTERS SERIES VALVE EXPANDED CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 27MECJ-502
Device Problems Fracture (1260); Material Split, Cut or Torn (4008)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/15/2019
Event Type  Injury  
Manufacturer Narrative
Investigation results will be provided in a subsequent submission.
 
Event Description
On (b)(6) 2019, a 27mm masters series valve expanded cuff was selected for implant in the mitral position.During positioning, a leaflet fractured.The valve was removed and all pieces were retrieved from the patient.The procedure and bypass were reported as extended a clinically significant time (15 minutes) while exchanging the valve and the patient remained hemodynamically stable throughout.Another 27mm masters valve was implanted successfully and post-operatively, the patient is reported to be stable.
 
Manufacturer Narrative
The reported event of a dislodged leaflet was confirmed.Both leaflets were dislodged from the orifice and returned with the valve.The butterfly apices were also chipped.No other damage was noted.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.The cause of the leaflet dislodgment could not be conclusively determined.Please note, per the instructions for use artmt100122073 ver.A, "using the valve holder/rotator and the flexible valve holder handle model 905-hh, or the rigid valve holder handle model 905-rhh, rotate the valve in situ to the desired position.The valve should rotate freely.If resistance is noted, the valve holder/rotator may not be properly seated in the valve, the valve may not be in the fully closed position, or the valve may be oversized.If the valve does not rotate freely, do not force valve rotation.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM MASTERS SERIES VALVE EXPANDED CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
MDR Report Key8464690
MDR Text Key140314526
Report Number2648612-2019-00022
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734006453
UDI-Public05414734006453
Combination Product (y/n)N
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/09/2023
Device Model Number27MECJ-502
Device Catalogue Number27MECJ-502
Device Lot Number6629353
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2019
Was the Report Sent to FDA? No
Date Manufacturer Received04/24/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-