It was reported that due to an image quality issue with an el-18-4 transducer model during a mammogram, a patient needed to be rescanned with a different transducer to ensure a proper diagnosis.Upon rescanning the patient, the customer identified the need for a breast biopsy.There was no injury associated with this event.
|
Evaluation of the returned el18-4 transducer confirmed the device had undergone physical damage.A thorough investigation of the device could not reproduce the reported image quality issue.However, inspection of the transducer revealed visible corrosion to the connector, and a 007 thermal circuit error was generated during diagnostic testing.The physical damage to the transducer inhibited the overall performance of the device and is indicative of improper maintenance.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|