ENCORE MEDICAL, L.P. TURON SHOULDER; TURON PEGGED GLENOID, SZ 50MM, EPLUS Back to Search Results
Model Number 521-01-250
Device Problems Appropriate Term/Code Not Available (3191); Material Split, Cut or Torn (4008)
Patient Problems Inflammation (1932); Swelling (2091)
Event Date 03/04/2019
Event Type  Injury  
Manufacturer Narrative
The reason for this revision surgery was due to inflammation and torn rotator cuff.The previous surgery and the surgery detailed in this investigation occurred 2.2 years apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr), shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to inflammation and torn rotator cuff.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event Description
Revision surgery - due to inflammation and extra fluid in the shoulder; torn rotator cuff.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURON SHOULDER
Type of Device
TURON PEGGED GLENOID, SZ 50MM, EPLUS
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758 6313
Manufacturer (Section G)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758 6313
Manufacturer Contact
teffany hutto
9800 metric blvd.
austin, TX 78758-6313
MDR Report Key8471508
MDR Text Key140586114
Report Number1644408-2019-00248
Device Sequence Number1
Product Code PAO
UDI-Device Identifier00888912145381
UDI-Public(01)00888912145381
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/01/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/28/2018
Device Model Number521-01-250
Device Catalogue Number521-01-250
Device Lot Number801N1006
Was Device Available for Evaluation? No
Date Manufacturer Received03/04/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/25/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age66 YR
-
-