• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TULSA DENTAL PRODUCTS LLC ASEPTICO ENDO DTC MOTOR; HANDPIECE, DIRECT DRIVE, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TULSA DENTAL PRODUCTS LLC ASEPTICO ENDO DTC MOTOR; HANDPIECE, DIRECT DRIVE, AC-POWERED Back to Search Results
Catalog Number AEU-25
Device Problems Mechanical Problem (1384); Retraction Problem (1536)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
While no serious injury resulted in this event, there has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Event Description
It was reporting that an aseptico dtc endo motor was improperly auto-reversing; no injury resulted.
 
Manufacturer Narrative
Found no torque, need u3 replaced.Both overlays, software needs updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASEPTICO ENDO DTC MOTOR
Type of Device
HANDPIECE, DIRECT DRIVE, AC-POWERED
Manufacturer (Section D)
TULSA DENTAL PRODUCTS LLC
608 rolling hills drive
johnson city TN 37604
MDR Report Key8471623
MDR Text Key140640127
Report Number2320721-2019-00056
Device Sequence Number1
Product Code EKX
Combination Product (y/n)N
PMA/PMN Number
K030163
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 05/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/01/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberAEU-25
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2019
Date Manufacturer Received04/24/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-