• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRIDE MOBILITY PRODUCTS PRIDE MOBILITY PRODUCTS; ELECTRIC POSITIONING CHAIR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRIDE MOBILITY PRODUCTS PRIDE MOBILITY PRODUCTS; ELECTRIC POSITIONING CHAIR Back to Search Results
Model Number LC525IM
Device Problem Fire (1245)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The customer discarded the hand control.The remainder of the device has not yet been made available for evaluation.Should further information or the device become available, a follow-up report will then be issued.
 
Event Description
Alleges being reclined back in the chair when she smelled something burning and alleges the hand control caught on fire.Alleges she yanked the control from the chair and threw it outside and roommate threw the hand control away.
 
Manufacturer Narrative
Only the lift chair frame and motor were returned for evaluation.As stated in the text, the hand control was disposed of by the consumer's roommate and there was no evidence of fire on the two components that were returned.
 
Event Description
Alleges being reclined back in the chair when she smelled something burning and alleges the hand control caught on fire.Alleges she yanked the control from the chair and threw it outside and roommate threw the hand control away.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIDE MOBILITY PRODUCTS
Type of Device
ELECTRIC POSITIONING CHAIR
Manufacturer (Section D)
PRIDE MOBILITY PRODUCTS
401 york ave
duryea PA 18642
Manufacturer (Section G)
N/A
n/a
n/a
n/a
Manufacturer Contact
kelly livingston
401 york ave
duryea, PA 18642
5706024056
MDR Report Key8473967
MDR Text Key140635721
Report Number2530130-2019-00034
Device Sequence Number1
Product Code INO
UDI-Device Identifier00606509400107
UDI-Public00606509400107
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091578
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/02/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberLC525IM
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/28/2019
Date Manufacturer Received03/28/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/29/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other;
-
-