Catalog Number 222140
Device Problem Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).The expiration date is unknown.
 
Event Description
It was reported by the sales rep via cst that during an unknown procedure, the doctor could not get any torque when attempting to insert the 4mm fastin anchor into the patient.The case was completed with a different type of implant with no patient harm or delay.The implant was discarded.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The complaint device is not being returned, it was discarded by the customer, therefore, unavailable for a physical evaluation.This complaint cannot be confirmed.No non-conformance were identified for this part (222140) - lot number (3908374) combination as per qlik query executed 04/12/2019.No further information regarding the technique or instruments used has been provided to determine a root cause for this failure.If any additional information is obtained, this complaint will be re-opened to capture that information.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key8475968
MDR Text Key145911554
Report Number1221934-2019-56741
Device Sequence Number1
Product Code JDR
UDI-Device Identifier10886705007509
UDI-Public10886705007509
Combination Product (y/n)N
PMA/PMN Number
K945203
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Type of Report Initial,Followup
Report Date 03/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/02/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2021
Device Catalogue Number222140
Device Lot Number3908374
Was Device Available for Evaluation? No
Date Manufacturer Received04/12/2019
Patient Sequence Number1
-
-