The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The product lot number was not provided, therefore, the manufacturing records could not be reviewed.
|
During preparation for a medical procedure using the penumbra system ace68 reperfusion catheter (ace68), the ace68 was found to be slightly kinked at the distal end upon removal from the packaging.The damage to the ace68 was found prior to use and, therefore, it was not used in the procedure.The procedure was completed using another catheter.
|