MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ENTERRA; INTESTINAL STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ENTERRA; INTESTINAL STIMULATOR Back to Search Results
Model Number 37800
Device Problem Migration or Expulsion of Device (1395)
Patient Problem Pain (1994)
Event Date 03/27/2019
Event Type  Injury  
Manufacturer Narrative
Analysis results were not available at the time of this report.An additional report will be sent once analysis is completed.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer (con) via a manufacturer representative (rep) regarding a patient who was implanted with a neurostimulator for gastric stimulation.It was reported that the device was moving in the pocket causing pain.There were no factors noted that could have led or contributed to the issue.The device was checked and was functioning properly.The patient was getting a revision and replacement on (b)(6) 2019.It was unknown if the issue was resolved at the time of the report.No further complications were reported or anticipated.
 
Manufacturer Narrative
Analysis of the ins ((b)(4)) found the ins was functionally okay; there were insignificant anomalies.The ins was subjected to a series of standard tests that include, but is not limited to, visual inspection, output and telemetry testing, and functional testing.The ins passed functional testing.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key8486505
MDR Text Key141022123
Report Number3004209178-2019-06775
Device Sequence Number1
Product Code LNQ
UDI-Device Identifier00643169614246
UDI-Public00643169614246
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 06/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/05/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/14/2020
Device Model Number37800
Device Catalogue Number37800
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2019
Date Manufacturer Received05/29/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/20/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age46 YR
Patient Weight61
-
-