BOSTON SCIENTIFIC CORPORATION SPEEDBAND SUPERVIEW SUPER 7; LIGATOR, HEMORRHOIDAL Back to Search Results
Model Number M00542251
Device Problem Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/13/2019
Event Type  malfunction  
Manufacturer Narrative
The complainant was unable to report the device lot number; therefore, the manufacture and expiration date is unknown.(b)(4).According to the complainant, the suspect device has been disposed and is not available for return.If any further relevant information is received, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that two speedband superview super 7 devices were used for a demonstration performed on (b)(6) 2019.According to the complainant, the first band would not deploy.Reportedly, a band appeared in the middle of the seven bands was deploying first and caused the bands not to deploy.A second speedband device was used; however, the suture had an extra knot which made the loop too small for the ligator head to drop through it.Additionally, there was no difficulty in setting up the device.This device was not used in the patient or procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPEEDBAND SUPERVIEW SUPER 7
Type of Device
LIGATOR, HEMORRHOIDAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key8491116
MDR Text Key141181713
Report Number3005099803-2019-01575
Device Sequence Number1
Product Code FHN
UDI-Device Identifier08714729201960
UDI-Public08714729201960
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
510K EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 04/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/08/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00542251
Device Catalogue Number4225
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/13/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-