• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number WCD 4000
Device Problems Failure to Charge (1085); Device Emits Odor (1425)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/01/2019
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of battery charger sn (b)(4) has been completed.The reported problem (does not charge battery pack) was confirmed.As received, the charger would not recognize a test battery pack.Upon evaluation, the q1 current controlling transistor was shorted.The cause of the inability to charge the battery packs is the shorted transistor.The root cause of the shorted transistor cannot be positively identified.No adverse event resulted from the defective battery charger.
 
Event Description
A us distributor contacted zoll to report that a patient's battery charger was not properly charging the battery packs and 'smelled like smoke'.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
katelynn mains
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key8496587
MDR Text Key141436833
Report Number3008642652-2019-02664
Device Sequence Number0
Product Code MVK
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 04/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/09/2019
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/14/2019
Date Manufacturer Received03/22/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/25/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-