• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT ENDURITY MRI; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT ENDURITY MRI; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM2172
Device Problem Device Damaged Prior to Use (2284)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/05/2019
Event Type  malfunction  
Event Description
During implant procedure, the physician noted that the device header was damaged.A new device was used and successfully implanted.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of ruined header was not confirmed; however,, a cosmetic anomaly was noted.The device was received in normal working condition.The epoxy header surface appeared to be uneven and rough to the touch.Electrical and mechanical tests indicated normal pacer functionality.As a result of this finding, abbott is performing further investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURITY MRI
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8496921
MDR Text Key141375147
Report Number2017865-2019-05221
Device Sequence Number1
Product Code LWP
UDI-Device Identifier05414734509602
UDI-Public05414734509602
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P140033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/09/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2020
Device Model NumberPM2172
Device Lot NumberP000064827
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/17/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/16/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/19/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-