Catalog Number H1-M |
Device Problem
Difficult to Open or Close (2921)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/05/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
The physician was using an hawkone to treat a 50-60mm moderately calcified lesion with 70% stenosis in the left mid / proximal common femoral artery.The vasculature was moderately tortuous and the vessel diameter was 6-7mm.A 6mm non medtronic sheath, a 0.14 non medtronic guidewire and a 4-7mm non medtronic embolic protection was used in the procedure.The ifu was followed.It was reported that the plunger in the nose cone stopped operating.Another hawkone was used to complete the procedure.No patient injury was reported.
|
|
Manufacturer Narrative
|
Additional information: the position of the cutter upon removal of the device is unknown.The device was safely removed from the patient.No deformation was noted in the cutter.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Device evaluation visual inspection: the hawkone was inspected and observed the torque shaft beneath the strain relief was bent at an approximate 135-degree angle.The cutter driver power switch was in the off position.The distal assembly was inspected and observed a bend to coiled housing approximately 1.4 cm from the cutter window.The cutter was positioned approximately 1.2cm distal the cutter window.Tissue/biological debris was observed at the distal rim of the cutter.Traces of dried blood was noted within the cutter window.Functional testing: the torque shaft was straightened by hand and the cutter driver was powered on.The thumb switch was retracted, and the cutter was pulled back into the cutter window as intended.No damages to the cutter assembly were noted.The thumb switch attempted to be advanced but could not go further than the original location.The dft was placed over the distal assembly and flushed with water.Not all of the observed biologics could be flushed.The thumb switch was attempted to be advanced and could not pass the cutter window.It is possible damage to torque shaft and/or biologics behind the cutter may have interfered with the ability to advance.Image review the customer provided 2 cine images from the procedure.Both images show the lesion within the vessel with the guidewire inserted.The first image shows the what appears to be the distal assembly of the hawkone inserted the vessel.The provided cine could not identify the with certainty the location of the suspected cutter heard.The second photo shows the vessel with a guidewire placed within the vessel, but no ancillary devices appear loaded over the wire.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|