Device evaluation; the device was removed from the return packaging.It was noted the torque shaft was bent at a 90-degree angle.Dried blood and biological material were in the distal assembly.Inspection of the cutter window revealed the drive shaft and cutter were not in the window.Fracture observed, and further inspection revealed the cutter was located approximately 3.0cm distal to the cutter window in the distal assembly.The distal housing was cut to remove the cutter.The cutter measure approximately 1.0cm.It was attached to a portion of the shaft.Inspection of the cutter revealed dents and damage.Examination of the fractured end revealed a coil protruding upward (analysis photo 9).The thumb switch was advanced, and the proximal fracture was observed, the coils appeared twisted.The fractured ends were consistent with a ductile fracture.Image review: one collage image was received, consisting of two cine images.The cine images appear to be before and after cines of the treated targeted vessel.The two photographs appear to be identical and show the debris removed from the targeted lesion.The hawkone was not included in the cine images.The collages show the successful treatment of the lesion.If information is provided in the future, a supplemental report will be issued.
|