SMITH & NEPHEW, INC. HEALICOIL RSB SA 5.5MM W/1 UT & 1 UB BLK; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
Catalog Number 72203708 |
Device Problems
Break (1069); Therapeutic or Diagnostic Output Failure (3023)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/21/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).
|
|
Event Description
|
It was reported that, during a rotator cuff repair procedure, when screwing in, the healicoil screw thread was unscrewed.The anchor broke in the operative field, but all pieces were removed.A back-up device was available to complete the procedure without significant delay.No patient injuries were disclosed.
|
|
Manufacturer Narrative
|
The reported 5.5mm healicoil rsb sa, intended for use in treatment, will not be returned for evaluation.Without the reported product a visual evaluation cannot be performed and the customers complaint cannot be confirmed.From the information provided, the anchor broke in the patient.An exact root cause cannot be determined with confidence; however, factors that could have contributed to the reported event include: the insertion site not prepared with the recommended smith & nephew instrumentation.Use of excessive force during insertion can cause failure of the suture anchor or insertion device.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.
|
|