• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V200 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V200 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number V200
Device Problems Loss of Power (1475); Self-Activation or Keying (1557)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2019.Date of report: 15april2019.A follow-up report will be submitted once the investigation has been complete.
 
Event Description
The customer reported that the machine is turning off automatically.There was no patient involvement.Event date not specified, estimate used.
 
Manufacturer Narrative
Date rec'd by mfr: 06may2019.The customer confirmed the reported system shut-down issue.The customer reseated the power supply cable connections.The customer kept the ventilator under observation.System as of now is working ok.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V200 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key8515434
MDR Text Key142053041
Report Number2031642-2019-02234
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
PMA/PMN Number
K102054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 03/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/15/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV200
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/23/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-