Results: the returned penumbra system jet 7 reperfusion catheter (jet7) was kinked at approximately 92.0 cm and 94.0 cm from the hub.Conclusions: evaluation of the returned jet7 confirmed a kinked device.If the packaging lid is not removed from the tray prior to removal of the jet7, the catheter will likely become kinked.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
During preparation for a medical procedure, the physician found that a penumbra system jet 7 reperfusion catheter (jet7) from a penumbra system jet7 kit was kinked at the distal end upon removal from the packaging.The damage to the jet7 was found prior to use and, therefore, it was not used in the procedure.The procedure was completed using another jet7.
|