• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US DXTEND METAGLENE; DELTA XTEND IMPLANTS : SHOULDER METAGLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US DXTEND METAGLENE; DELTA XTEND IMPLANTS : SHOULDER METAGLENE Back to Search Results
Model Number 130760000
Device Problem Device Dislodged or Dislocated (2923)
Patient Problems Edema (1820); Fatigue (1849); Inflammation (1932); Muscular Rigidity (1968); Pain (1994); Loss of Range of Motion (2032); Tissue Damage (2104); Weakness (2145); Joint Disorder (2373); No Code Available (3191); Unspecified Tissue Injury (4559); Swelling/ Edema (4577)
Event Date 06/26/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot: null.Device history batch: null.Device history review: null.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The patient was revised to address loosening of the glenosphere at the bone to implant interface.The glenosphere broke off at mid screw of the metaglene.Relevant components retrieved and re implanted with new implants.Original stem and epiphysis were stable and not affected by the broken glenosphere.Right shoulder.
 
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : the device associated with this report was not returned to depuy synthes for evaluation.However a disc with x-rays was returned for investigation.The x-ray review found the metaglane disassociated from the glenosphere middle screw, confirming the reported allegation.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND METAGLENE
Type of Device
DELTA XTEND IMPLANTS : SHOULDER METAGLENE
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
b.p. 256
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st anthonys road
b.p. 256
leeds 69801
FR   69801
Manufacturer Contact
kara ditty-bovard
1210 ward avenue
west chester, PA 19380-0988
6107428552
MDR Report Key8518065
MDR Text Key142051328
Report Number1818910-2019-90749
Device Sequence Number1
Product Code HSD
UDI-Device Identifier10603295027737
UDI-Public10603295027737
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K192855
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 06/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/16/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number130760000
Device Catalogue Number130760000
Device Lot Number5257260
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/30/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age68 YR
Patient SexMale
-
-