• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH UNKNOWN ACCOLADE II SIZE 5 127°; HIP IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH UNKNOWN ACCOLADE II SIZE 5 127°; HIP IMPLANT Back to Search Results
Catalog Number UNK_SHC
Device Problem Material Deformation (2976)
Patient Problem Injury (2348)
Event Date 03/19/2019
Event Type  Injury  
Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event Description
The patient presented with aseptic loosening of the acetabular cup, which was revised in (b)(6) 2018.The femoral stem was left in place during the first revision.The cup loosened again, requiring a second revision.When we exposed the components, we noticed that the mdm liner/acetabular shell wore a notch into the neck of the femoral stem.We removed all components and inserted a restoration modular stem, multihole acetabular cup and constrained liner.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN ACCOLADE II SIZE 5 127°
Type of Device
HIP IMPLANT
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
Manufacturer Contact
collin neitzel
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key8518281
MDR Text Key142060086
Report Number0002249697-2019-01622
Device Sequence Number1
Product Code MEH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial
Report Date 04/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/16/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNK_SHC
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/19/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age63 YR
-
-