Catalog Number 295053-001 |
Device Problem
Connection Problem (2900)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/28/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to the patient because it does not prevent the freedom driver from performing its life-sustaining functions.The freedom driver has a redundant power source of onboard batteries.The freedom car charger will be returned to syncardia for evaluation.The results of the evaluation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The freedom car charger is component that enables the freedom driver to be plugged into a 12v vehicle power outlet.The customer, a syncardia certified hospital, reported that the patient's freedom car charger will not stay plugged into the freedom driver.
|
|
Manufacturer Narrative
|
The customer-reported issue of the car charger not staying plugged into patient's freedom driver was confirmed via the functional evaluation and additional testing.The connector was found not to slide freely and therefore did not lock into place as designed into the driver.The root cause cannot be confirmed but is likely material fatigue from prolonged usage.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4) follow-up report 1.
|
|