The reported event was confirmed as a manufacturing related issue due to a hole observed within 0.5 inches of the bifurcation point.The visual evaluation of the returned sample noted one opened (without original packaging), cut portion of the inlet tube and silicone catheter.The catheter balloon was inflated with 10 ml methylene blue solution (3 drops 1% aq methylene blue per 100ml distilled water) and immediately leaked from a pinhole (0.017") over the inflation lumen 0.4845" from the bifurcation.This was a manufacturing related issue in violation of inspection procedure, which states the "shaft shall be free of kinks, cuts, tears and irregular surfaces".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "(5) do not hold the device with forceps, etc.Avoid contact with any blades or sharp-edged instruments.[catheter damage may cause balloon rupture and accidental balloon removal or failure to deflate or remove the balloon.]" h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|