As reported, during the peripheral angioplasty, the powerflex pro pta (6mm15cm 135) balloon catheter did not enter trough the guidewire, there was much resistance; therefore, another balloon was used and worked well.There was no reported patient injury.A 6f vista brite tip catheter sheath introducer was used in the procedure.The device was stored and handled per the instructions for use ifu.There was no damage noted to the product packaging upon inspection prior to use.There was no reported difficulty removing the product from the packaging.There were no kinks noted on the device prior to use.The device was prepped according to instructions for use (ifu).The catheter was flushed with heparinized saline.The insertion difficulty was caused by a blockage of possibly injectable material.There was no excessive force applied on the device while inserting it over the guide wire.Other products were successfully used with the device prior to the encountered resistance.The catheter was not torqued against resistance.There was no unusual force used at any time during the procedure.The difficulty require that only the reported product was removed.The device was removed intact (in one piece) from the patient.The procedure was completed with another balloon.The procedure was completed successfully without patient injury.There are no procedural films available.The device will not be returned for analysis.
|