• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number ASKU
Device Problem Use of Incorrect Control/Treatment Settings (1126)
Patient Problem Corneal Edema (1791)
Event Date 03/31/2019
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported during a multiple surgical procedure when using a balanced phacoemulsification (phaco) tip there was no setting "seven" on the system and the staff was not able to add that setting.The physician tried setting "four" with the balanced tip and it made a funny noise and did not work at all.The tip was changed to another phaco tip and setting "four" was tried again which resulted in no phaco power.The surgeon decided to try setting "one" and increased the vacuum which helped.The setting issue caused an hour delay in the procedure and the patient experienced corneal edema which made the stents harder to insert.Additional information has been requested but not received.
 
Manufacturer Narrative
No further information was able to be obtained from this customer.With no additional, related information provided, the customers reported event was not able to be confirmed.The root cause cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key8535888
MDR Text Key142753986
Report Number2028159-2019-00717
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K120912
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/22/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberASKU
Device Catalogue NumberASKU
Was Device Available for Evaluation? No
Date Manufacturer Received05/17/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
HEALON GV; MYALUGIN RING; PROVISC; VISCOAT; VISION BLUE
Patient Age95 YR
-
-