|
Model Number J172 |
Device Problem
Defective Device (2588)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/11/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Following completion of analysis, a follow-up report will be completed.
|
|
Event Description
|
It was reported that during the attempted implant of this device, upon lead insertion, the pacing frequency was lower than the marked frequency.For this reason, the device was removed and returned for analysis.No resulting adverse patient effects.
|
|
Manufacturer Narrative
|
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.A visual inspection of the device header and case noted no anomalies.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
|
|
Event Description
|
It was reported that during the attempted implant of this device, upon lead insertion, the pacing frequency was lower than the marked frequency.For this reason, the device was removed and returned for analysis.No resulting adverse patient effects.
|
|
Search Alerts/Recalls
|
|
|