SMITH & NEPHEW, INC. LIGHT SOURCE, 500XL, XENON; IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
|
Back to Search Results |
|
Catalog Number 72200568 |
Device Problem
Device Emits Odor (1425)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/11/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).
|
|
Event Description
|
It was reported that at the start of the case a burning smell was coming from the light source unit.The resistor on the power supply was blown.The procedure was completed with a backup device with no significant delay or patient injury reported.
|
|
Manufacturer Narrative
|
The reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was a relationship found between the subject device and the reported incident.A service assessment of the unit revealed that the fan was not working.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.The complaint was confirmed and the root cause has been associated with electrical component failure.Factors, unrelated to the manufacturing and design of the device that could have contributed to the reported event, include a defective fan.Please note the instructions for use provided with the device caution against installing the device in an environment in which the vents and fans are obstructed.
|
|