TOSOH BIOSCIENCE, INC. TOSOH HLC-723G8 ANALYZER G8 Back to Search Results
Model Number G8
Device Problem Decrease in Pressure (1490)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The customer evaluated the analyzer with the assistance of field service engineer (fse) and determined that air was entering pump due to buffer levels running low.The fse instructed the customer to prime all buffers.The analyzer was operating as expected.There was no further action required by fse.
 
Event Description
This report summarizes 1 malfunction event on the g8 analyzer.The review of the event indicated that the g8 analyzer experienced pressure error messages, which caused delayed reporting of critical patient results.This report was received from one source.There was no indication of patient intervention or adverse health consequences due to the delayed reporting in patient results.This event did not necessitate remedial action to prevent an unreasonable risk of substantial harm to the public health.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH BIOSCIENCE, INC.
3600 gantz road
grove city OH 43123
Manufacturer (Section G)
TOSOH BIOSCIENCE, INC.
3600 gantz road
grove city OH 43123
Manufacturer Contact
doria esquivel
6000 shoreline court
ste 101
south san francisco, CA 94080
6506368123
MDR Report Key8543677
MDR Text Key147062227
Report Number3005529799-2019-00086
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
PMA/PMN Number
K131580
Number of Events Reported1
Summary Report (Y/N)Y
Report Source Manufacturer
Source Type health professional
Type of Report Initial
Report Date 04/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-