BAXTER HEALTHCARE - IRVINE FOLFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C4711K
Device Problem Material Integrity Problem (2978)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The single-use device was received for evaluation.Visual inspection revealed fluid inside the bag that contained the device.When the device was removed from the bag, the cause of the fluid was found to be an untightened blue winged luer cap.A functional leak test was performed by filling the device with water and manually tightening the blue winged luer cap.During and after fill, no signs of a leak were observed from the device.The reported condition was not verified.The device was found to be conforming product.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
This report summarizes 1 malfunction events.It was reported that a small volume folfusor had an unspecified issue during an unspecified process step.It was not specified whether there was patient involvement.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE - IRVINE
irvine CA
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key8550228
MDR Text Key143931053
Report Number1416980-2019-02263
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412082172
UDI-Public(01)00085412082172
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)Y
Report Source Manufacturer
Type of Report Initial
Report Date 04/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/25/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator No Information
Device Catalogue Number2C4711K
Device Lot Number18J042
Date Manufacturer Received03/31/2019
Type of Device Usage N
Patient Sequence Number1
-
-