|
Model Number N/A |
Device Problem
No Apparent Adverse Event (3189)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/27/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Siemens went on site and performed a total service call.The service engineer ran the dry test and observed all rlus in the low 30 range.The acid/base reservoirs were decontaminated.The sample was run as a single neat sample, auto-diluted x10 and all results were >600 u/ml.On a second visit, a total service call was performed.The acid/base reservoirs and debubblers and tubings were replaced.The acid/base lines, pumps, valves and reservoirs were decontaminated.A leaking ancillary diluter syringe was found and replaced.The pmt, sample digital diluter, sample syringe and reagent probe 3 syringe were replaced proactively.All probe alignments were check.The vacuum pressure was checked and the system was checked for leaks.Dry, wetwater and wetwash1 tests were performed and all were acceptable.All assays were calibrated and qc was run.The cause for the discordant ca 125ii result is unknown.Siemens is investigating.The dilutions section of the instructions for use (ifu) states: "serum samples with ca 125 levels greater than 600 u/ml must be diluted and retested to obtain accurate results." the interpretation of results section of the ifu states: "results of this assay should always be interpreted in conjunction with patient's medical history, clinical presentation and other findings." the limitations section of the ifu states: "note: do not interpret levels of ca 125 as absolute evidence of the presence or the absence of malignant disease.Before treatment, patients with confirmed ovarian carcinoma frequently have levels of ca 125 within the range observed in healthy individuals.Elevated levels of ca 125 can be observed in patients with nonmalignant diseases.Measurements of ca 125 should always be used in conjunction with other diagnostic procedures, including information from the patient's clinical evaluation.The concentration of ca 125 in a given specimen determined with assays from different manufacturers can vary due to differences in assay methods, calibration, and reagent specificity.Ca 125 determined with different manufacturers' assays will vary depending on the method of standardization and antibody specificity.The concentration of ca 125 in a given specimen determined with assays from different manufacturers can vary because of differences in assay methods, calibration, and reagent specificity.Ca 125 determined with different manufacturers' assays will vary depending on the method of standardization and antibody specificity.Therefore, it is important to use assay specific values to evaluate quality control results.".
|
|
Event Description
|
Customer observed a neat sample result for advia centaur xp ca 125ii that was low compared to diluted values.The physician questioned the results.There are no reports that treatment was altered or prescribed or adverse health consequences due to the low advia centaur xp ca 125ii result.
|
|
Manufacturer Narrative
|
Mdr was filed on april 25, 2019 reporting the customer observation of a neat sample result that was low compared to diluted values on the advia centaur xp ca 125ii assay.May 17, 2019 - additional information: a review of post service testing indicates that the instrument is fully functional.Siemens received the following information from the customer regarding the patient: the draw dates of the samples are (b)(6) 2019 and (b)(6) 2019.The samples were drawn into greiner sst tubes to a volume of 4 ml.The samples were poured off samples using a 13/75 tube.Advia centaur ca 125ii reagent lot: 181 and calibrator 15 lot c1533 was used for both samples on for both samples on advia centaur xp irl24811127.Historical ca125 data for the patient.(b)(6) 19 = 776 (corrected from 470), (b)(6) 19 = 962 (corrected from 570), (b)(6) 19 = 1171, (b)(6) 19 = 1569, (b)(6) 18 = 1534.A comprehensive metabolic panel was performed on both samples and the results were as expected.The patient is currently taking the following medications; ascorbic acid (vitamin c), b complex vitamins, calcium carbonate (vitamin d3), cholecalciferol, vitamin d3, clindamycin (cleocin), cyanocobalamin, diazepam (valium), folic acid (folvite), glucosamine sulfate, heparin, magnesium oral, melatonin oral, prochlorperazine (compazine), propylene glycol, tocopheryl acetate, vitamin k2, zinc oral.Siemens continues to investigate.
|
|
Manufacturer Narrative
|
Mdr (b)(4) was filed on (b)(6)2019 reporting the customer observation of a neat sample result that was low compared to diluted values on the advia centaur xp ca 125ii assay.Mdr (b)(4) supplemental report 1 was file on (b)(6)2019 for additional information.(b)(6)2019 - additional information: siemens had inquired and was informed that the patient had not undergone retinal fluorescein angiography.(b)(6)2019 - additional information: based on review of the information provided, the cause of the low advia centaur xp ca 125ii neat result compared to diluted values is unknown.Quality control samples were in range at the time of the event.There were no issues observed by the customer with other patient samples, indicating that the instrument and reagents were performing acceptably.Sample integrity or other preanalytical variables cannot be ruled out as potential contributing factors.Specific testing was not identified regarding why the discordant sample would read on the calibration curve, then recover higher after repeat testing on later dates.A systemic product non-conformance has not been identified.The instrument is performing within specifications.No further evaluation of the device is required.Section h6 result code and conclusion code were updated to reflect the additional information.
|
|
Search Alerts/Recalls
|
|
|