SMITH & NEPHEW ORTHOPAEDICS LTD BHR ACETABULAR CUP 54MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
|
Back to Search Results |
|
Catalog Number 74122154 |
Device Problems
Break (1069); Improper or Incorrect Procedure or Method (2017)
|
Patient Problems
Injury (2348); No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/28/2019 |
Event Type
malfunction
|
Event Description
|
Wire holding the cup to the impactor broke when positioning the cup into the acetabulum, before impaction.As a result the cup could not be implanted.Surgeon reamed up to 56mm and twice attempted to use 74122156.As a result of the increased reaming it was found that the posterior wall of the acetabulum had become too thin to hold the bhr cup, at this stage the only option remaining was to convert to a total hip replacement using an acetabular cup with additional screw fixation.48mm head had already been opened.Replacement of 1x 74122156 urgently requested.Other items were from loan and customer services have already been made aware.
|
|
Manufacturer Narrative
|
Smith + nephew is submitting this report pursuant to the provisions of 21cfr, part 803.This report may be based upon information which smith + nephew has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, smith + nephew, or its employees, that the report constitutes an admission that the device, smith + nephew or its employees caused or contributed to the potential event described in this report.
|
|