• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD BHR ACETABULAR CUP 54MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD BHR ACETABULAR CUP 54MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74122154
Device Problems Break (1069); Improper or Incorrect Procedure or Method (2017)
Patient Problems Injury (2348); No Known Impact Or Consequence To Patient (2692)
Event Date 03/28/2019
Event Type  malfunction  
Event Description
Wire holding the cup to the impactor broke when positioning the cup into the acetabulum, before impaction.As a result the cup could not be implanted.Surgeon reamed up to 56mm and twice attempted to use 74122156.As a result of the increased reaming it was found that the posterior wall of the acetabulum had become too thin to hold the bhr cup, at this stage the only option remaining was to convert to a total hip replacement using an acetabular cup with additional screw fixation.48mm head had already been opened.Replacement of 1x 74122156 urgently requested.Other items were from loan and customer services have already been made aware.
 
Manufacturer Narrative
Smith + nephew is submitting this report pursuant to the provisions of 21cfr, part 803.This report may be based upon information which smith + nephew has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, smith + nephew, or its employees, that the report constitutes an admission that the device, smith + nephew or its employees caused or contributed to the potential event described in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BHR ACETABULAR CUP 54MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa CV31 3HL
UK  CV31 3HL
MDR Report Key8553906
MDR Text Key143446752
Report Number3005975929-2019-00182
Device Sequence Number1
Product Code NXT
UDI-Device Identifier03596010552297
UDI-Public03596010552297
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/26/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/27/2023
Device Catalogue Number74122154
Device Lot Number18BW19013
Was Device Available for Evaluation? No
Date Manufacturer Received03/28/2019
Patient Sequence Number1
-
-