• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - DOMINICAN REPUBLIC INTRAVIA CONTAINER EMPTY; CONTAINER, I.V.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - DOMINICAN REPUBLIC INTRAVIA CONTAINER EMPTY; CONTAINER, I.V. Back to Search Results
Catalog Number 2B8011
Device Problem Failure to Disconnect (2541)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
One actual sample was received for evaluation.The returned unit was labeled for single use.A visual inspection was performed and it was noticed that the bag had attached a spike.Functional testing was performed, when we tried to remove the spike from the bag, the membrane tube came out of the bag, it was impossible remove the spike.The reported condition was verified for the returned unit.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
This report summarizes 2 malfunction events.It was reported that an unspecified amount of 150 ml bags were difficult to remove from iv sets at the bag port.This event occurred during use of the device.These events involved patients with no patient consequences.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRAVIA CONTAINER EMPTY
Type of Device
CONTAINER, I.V.
Manufacturer (Section D)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
haina, san cristobal
Manufacturer (Section G)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
carretera sanchez km 18.5
parque industrial itabo, piisa
haina san cristobal
DR  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key8555326
MDR Text Key144988597
Report Number1416980-2019-02312
Device Sequence Number1
Product Code KPE
UDI-Device Identifier00085412003467
UDI-Public(01)00085412003467
Combination Product (y/n)N
Number of Events Reported2
Summary Report (Y/N)Y
Report Source Manufacturer
Type of Report Initial
Report Date 04/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/26/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator No Information
Device Catalogue Number2B8011
Device Lot NumberDR18G24070
Date Manufacturer Received03/31/2019
Type of Device Usage N
Patient Sequence Number1
-
-