• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND 9616671 ACTIS BROACH SZ 4; BROACHES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND 9616671 ACTIS BROACH SZ 4; BROACHES Back to Search Results
Catalog Number 201001040
Device Problems Device-Device Incompatibility (2919); Material Deformation (2976); Mechanical Jam (2983); Physical Resistance/Sticking (4012)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/08/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The broaches are getting stuck in the me 1000 adapter.As per sales rep, he spoke with (b)(6) (event reporter), the reason that the broaches are getting stuck is a result of the broach trunnion being damaged while using the calcar planer while not fully seated over the broach trunnion.This is causing burrs on the broach trunnions.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot null.Device history batch null.Device history review null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIS BROACH SZ 4
Type of Device
BROACHES
Manufacturer (Section D)
DEPUY IRELAND 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key8560913
MDR Text Key143433834
Report Number1818910-2019-91857
Device Sequence Number1
Product Code KWL
UDI-Device Identifier10603295393665
UDI-Public10603295393665
Combination Product (y/n)N
PMA/PMN Number
K150862
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number201001040
Device Lot NumberPG276749
Was Device Available for Evaluation? No
Date Manufacturer Received05/31/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-