Model Number MMT-723NAB
Device Problem Difficult or Delayed Activation (2577)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/23/2018
Event Type  malfunction  
Manufacturer Narrative
No button error alarm noted during testing.Unit had unresponsive down button due to unlocked j2/lcd keypad connector.The asr exemption e2015043 was revoked on february 4, 2019.This event was previously reported as an asr.Additional information or analysis results regarding this event will be reported as an mdr.
 
Event Description
The customer reported via phone call the insulin pump had a keypad anomaly, non of the buttons are responding.The customer¿s blood glucose was 339mg/dl at the time of incident, declined to troubleshoot for high blood glucose levels.Customer stated that the insulin pump.No significant events leading to keypad anomaly were observed.Keypad anomaly troubleshooting was performed and confirmed that time is advancing.The customer was advised to discontinue use of pump and revert to back-up plan.The insulin pump is being replaced and is expected to return for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAB
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key8568073
MDR Text Key143684529
Report Number3004209178-2019-90877
Device Sequence Number1
Product Code OYC
UDI-Device Identifier00643169507210
UDI-Public(01)00643169507210
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 05/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/01/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-723NAB
Device Catalogue NumberMMT-723NAB
Device Lot NumberB5723NABJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2019
Date Manufacturer Received04/29/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age53 YR
Patient Weight152
-
-