Results: during functional testing, a pipe cleaner was inserted into the pump vacuum inlet and blood was observed inside the vacuum tubing.Conclusions: evaluation of the returned pump max confirmed blood was aspirated inside the vacuum pump.The complaint reported that the aspiration tubing was connected directly to the pump max vacuum inlet, rather than the canister, supplied by penumbra.If fluid is aspirated into the vacuum assembly, the pump may not function properly.Penumbra pumps are visually inspected and functionally tested during incoming quality inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
|